20240531193620_副本 Operation video

 

Calcium and magnesium concentration in water and total hardness of water quality test kit

Scope of Application This reagent kit can be widely used for the determination of calcium concentration/magnesium concentration/total hardness of water in drinking water, source water, and aquaculture water.

Measurement Range Titration determination with calcium concentration above 20mg/L, magnesium concentration above 12mg/L, and total hardness above 50mg/L.

Calcium determination Take a water sample to a 5mL mark in a test tube, add 3 drops of calcium magnesium reagent (1), shake well, add 1 small spoonful of calcium magnesium reagent (2), shake well (the water sample turns red at this time), add calcium magnesium reagent (5) dropwise, shake while adding, and drop until the water color changes from red to blue. Record the number of drops of calcium magnesium reagent (5) and set it to A.

Hardness Measurement Rinse the test tube several times with water sample, then take the water sample to the 5mL mark, add 3 drops of calcium magnesium reagent (3), shake well, add 1 small spoonful of calcium magnesium reagent (4), shake well (at this time, the water sample turns red), add calcium magnesium reagent (5) to the water sample, shake while adding, and drop until the water color changes from red to blue. Record the number of drops of calcium magnesium reagent (5) and set it to B.

Result Calculation

Calcium concentration (mg/L)=A × 20

Magnesium concentration (mg/L)=(B-A) x 12

Total hardness (measured in calcium carbonate), Mg/L)=50 × B

Precautions

1This method is a rapid on-site detection method, and accurate quantification should be based on the national standard detection method.

2Once the reagent splashes into the eyes, rinse with plenty of water.

Kit Configuration 5ml/bottle for reagent 1 and 3, 0.5g/bottle for reagent 2 and 4, 10ml/2 bottle for reagent 5, 1 medicine spoon, and 1 test tube. Based on a total hardness of 450mg/L, the reagent can be used for 22 measurements. Calculated at 200mg/L, it can be used for 50 measurements.

Reagent Storage Store in a cool, dry place and out of reach of children.

 

 

0000

001_副本

Produced by Beijing Shian Xunda Technology Co., Ltd., China      

E-mailbjzw@vip.163.com